Detail publikace

Electro-mechanical Lung Simulator Using Polymer and Organic Human Lung Equivalents for Realistic Breathing Simulation

PAŠTĚKA, R. FORJAN, M. SAUERMANN, S. DRAUSCHKE, A.

Originální název

Electro-mechanical Lung Simulator Using Polymer and Organic Human Lung Equivalents for Realistic Breathing Simulation

Typ

článek v časopise ve Web of Science, Jimp

Jazyk

angličtina

Originální abstrakt

Simulation models in respiratory research are increasingly used for medical product development and testing, especially because in-vivo models are coupled with a high degree of complexity and ethical concerns. This work introduces a respiratory simulation system, which is bridging the gap between the complex, real anatomical environment and the safe, cost-effective simulation methods. The presented electro-mechanical lung simulator, xPULM, combines in-silico, ex-vivo and mechanical respiratory approaches by realistically replicating an actively breathing human lung. The reproducibility of sinusoidal breathing simulations with xPULM was verified for selected breathing frequencies (10-18 bpm) and tidal volumes (400-600 ml) physiologically occurring during human breathing at rest. Human lung anatomy was modelled using latex bags and primed porcine lungs. High reproducibility of flow and pressure characteristics was shown by evaluating breathing cycles (n(Total) = 3273) with highest standard deviation vertical bar 3 sigma vertical bar for both, simplified lung equivalents (mu(V) = 23.98 +/- 1.04 l/min, mu(P) = -0.78 +/- 0.63 hPa) and primed porcine lungs (mu(V) = 18.87 +/- 2.49 l/min, mu(P) = -21.13 +/- 1.47 hPa). The adaptability of the breathing simulation parameters, coupled with the use of porcine lungs salvaged from a slaughterhouse process, represents an advancement towards anatomically and physiologically realistic modelling of human respiration.

Klíčová slova

respiration simulation; biomedical electromechanical; systems, alternative to animal testing; lung simulation; primed porcine lungs; biomedical engineering education;

Autoři

PAŠTĚKA, R.; FORJAN, M.; SAUERMANN, S.; DRAUSCHKE, A.

Vydáno

24. 12. 2019

Nakladatel

NATURE PUBLISHING GROUP

Místo

LONDON

ISSN

2045-2322

Periodikum

Scientific Reports

Ročník

9

Číslo

9

Stát

Spojené království Velké Británie a Severního Irska

Strany od

1

Strany do

12

Strany počet

12

URL

Plný text v Digitální knihovně

BibTex

@article{BUT162245,
  author="Richard {Paštěka} and Mathias {Forjan} and Stefan {Sauermann} and Andreas {Drauschke}",
  title="Electro-mechanical Lung Simulator Using Polymer and Organic Human Lung Equivalents for Realistic Breathing Simulation",
  journal="Scientific Reports",
  year="2019",
  volume="9",
  number="9",
  pages="1--12",
  doi="10.1038/s41598-019-56176-6",
  issn="2045-2322",
  url="https://www.nature.com/articles/s41598-019-56176-6#article-info"
}