Detail publikace

Electrical impedance reconstruction measured by modified electrical impedance spectrometry

Originální název

Electrical impedance reconstruction measured by modified electrical impedance spectrometry

Anglický název

Electrical impedance reconstruction measured by modified electrical impedance spectrometry

Jazyk

en

Originální abstrakt

Electrical impedance characterizes the properties of an material when passing through an alternating electric current. The paper describes reconstruction of electrical impedance measured by multi-electrode resistive method of impedance spectrometry extended to electrical impedance tomography with Z-meter IV device in laboratory conditions. The measurement was carried out on the physical model of cylinder with a drinking water. To this environment was placed an asymmetrically glass, which represented an anomaly in the environment. The Z-meter device was used to measuring the components (real and imaginary) of electrical impedance of a given environment at a frequency of 8 kHz. Then values of the electrical impedance were plotted to the Surfer program, where reconstruction of the electrical impedance was carried out by means of triangulation with linear interpolation.

Anglický abstrakt

Electrical impedance characterizes the properties of an material when passing through an alternating electric current. The paper describes reconstruction of electrical impedance measured by multi-electrode resistive method of impedance spectrometry extended to electrical impedance tomography with Z-meter IV device in laboratory conditions. The measurement was carried out on the physical model of cylinder with a drinking water. To this environment was placed an asymmetrically glass, which represented an anomaly in the environment. The Z-meter device was used to measuring the components (real and imaginary) of electrical impedance of a given environment at a frequency of 8 kHz. Then values of the electrical impedance were plotted to the Surfer program, where reconstruction of the electrical impedance was carried out by means of triangulation with linear interpolation.

BibTex


@inproceedings{BUT159750,
  author="Zuzana {Műnsterová} and Jana {Pařílková} and Jan {Dušek} and Jan {Mikulka}",
  title="Electrical impedance reconstruction measured by modified electrical impedance spectrometry",
  annote="Electrical impedance characterizes the properties of an material when passing through an alternating electric current. The paper describes reconstruction of electrical impedance measured by multi-electrode resistive method of impedance spectrometry extended to electrical impedance tomography with Z-meter IV device in laboratory conditions. The measurement was carried out on the physical model of cylinder with a drinking water. To this environment was placed an asymmetrically glass, which represented an anomaly in the environment. The Z-meter device was used to measuring the components (real and imaginary) of electrical impedance of a given environment at a frequency of 8 kHz. Then values of the electrical impedance were plotted to the Surfer program, where reconstruction of the electrical impedance was carried out by means of triangulation with linear interpolation.",
  address="Brno University of Technology, VUTIUM",
  booktitle="EUREKA 2019 - 7th Colloquium and Working Session",
  chapter="159750",
  edition="neuv.",
  howpublished="print",
  institution="Brno University of Technology, VUTIUM",
  number="1",
  year="2019",
  month="october",
  pages="93--96",
  publisher="Brno University of Technology, VUTIUM",
  type="conference paper"
}