Detail publikace

Computational Modelling of the Shape Deviations of the Sphere Surfaces of Ceramic Heads of Hip Joint Replacement

FUIS, V. KOUKAL, M. FLORIAN, Z. JANÍČEK, P.

Originální název

Computational Modelling of the Shape Deviations of the Sphere Surfaces of Ceramic Heads of Hip Joint Replacement

Anglický název

Computational Modelling of the Shape Deviations of the Sphere Surfaces of Ceramic Heads of Hip Joint Replacement

Jazyk

en

Originální abstrakt

The problem of the damage of the bioimplants is very important today. In the endoprosthesis surgery there is a large percentage of implant defects, which cause the failure of the whole prosthesis. One kind of the total hip replacement functionality loss is acetabular cup pull-off from pelvis bone. This paper is aimed at manufacture perturbations (shape deviations) analysis as one of the possible reasons to this kind of failure. Dimension and of geometry manufacturing perturbations (roundness) were analyzed in detail. It was found, that these perturbations affect considered values of contact pressure and frictional moment. Contact pressure and frictional moment are quantities affecting replacement success and durability.

Anglický abstrakt

The problem of the damage of the bioimplants is very important today. In the endoprosthesis surgery there is a large percentage of implant defects, which cause the failure of the whole prosthesis. One kind of the total hip replacement functionality loss is acetabular cup pull-off from pelvis bone. This paper is aimed at manufacture perturbations (shape deviations) analysis as one of the possible reasons to this kind of failure. Dimension and of geometry manufacturing perturbations (roundness) were analyzed in detail. It was found, that these perturbations affect considered values of contact pressure and frictional moment. Contact pressure and frictional moment are quantities affecting replacement success and durability.

Dokumenty

BibTex


@inproceedings{BUT104713,
  author="Vladimír {Fuis} and Milan {Koukal} and Zdeněk {Florian} and Přemysl {Janíček}",
  title="Computational Modelling of the Shape Deviations of the Sphere Surfaces of Ceramic Heads of Hip Joint Replacement",
  annote="The problem of the damage of the bioimplants is very important today. In the endoprosthesis surgery there is a large percentage of implant defects, which cause the failure of the whole prosthesis. One kind of the total hip replacement functionality loss is acetabular cup pull-off from pelvis bone. This paper is aimed at manufacture perturbations (shape deviations) analysis as one of the possible reasons to this kind of failure. Dimension and of geometry manufacturing perturbations (roundness) were analyzed in detail. It was found, that these perturbations affect considered values of contact pressure and frictional moment. Contact pressure and frictional moment are quantities affecting replacement success and durability.",
  address="Springer",
  booktitle="XI Mediterranean Conference on Medical and Biological Engineering and Computing 2013",
  chapter="104713",
  doi="10.1007/978-3-319-00846-2_22",
  edition="IFMBE Proceedings",
  howpublished="electronic, physical medium",
  institution="Springer",
  year="2013",
  month="september",
  pages="89--92",
  publisher="Springer",
  type="conference paper"
}