Detail publikace

Measurement of concentration of air ions in the laboratories and cave

ROUBAL, Z. SZABÓ, Z. BARTUŠEK, K. STEINBAUER, M.

Originální název

Measurement of concentration of air ions in the laboratories and cave

Český název

Measurement of concentration of air ions in the laboratories and cave

Typ

článek ve sborníku ve WoS nebo Scopus

Jazyk

cs

Originální abstrakt

A gerdien tube (GT) is instrumental towards the measurement of air ions concentration assuming that the volume of tested air has not been limited and the ions are continuously generated. In the time domain, it is possible to measure only ions of one polarity by means of a gerdien tube; then, following a certain interval, ions of the other polarity can be measured. Ionized air is sucked into the gerdien tube by a fan. A homogeneous electric field is set between the inner and the outer electrode. If an electric ion shows a negative electric charge and the collecting inner electrode has a positive electric potential, the ion is progressively attracted to the inner electrode. Provided that the ion impinges upon the electrode, it will induce an electric current that is measured by the help of a sensitive electrometric picoampermeter. The velocity of the ions motion in the electric field can be described by mobility k [ m2*V-1*s-1].

Český abstrakt

A gerdien tube (GT) is instrumental towards the measurement of air ions concentration assuming that the volume of tested air has not been limited and the ions are continuously generated. In the time domain, it is possible to measure only ions of one polarity by means of a gerdien tube; then, following a certain interval, ions of the other polarity can be measured. Ionized air is sucked into the gerdien tube by a fan. A homogeneous electric field is set between the inner and the outer electrode. If an electric ion shows a negative electric charge and the collecting inner electrode has a positive electric potential, the ion is progressively attracted to the inner electrode. Provided that the ion impinges upon the electrode, it will induce an electric current that is measured by the help of a sensitive electrometric picoampermeter. The velocity of the ions motion in the electric field can be described by mobility k [ m2*V-1*s-1].

Klíčová slova

air ions, Gerdien tube

Rok RIV

2013

Vydáno

11.09.2013

Místo

Brno, Czech republic

ISBN

978-80-214-4759-2

Kniha

International Interdisciplinary PhD Workshop 2013, Proceedings

Strany od

182

Strany do

186

Strany počet

5

BibTex


@inproceedings{BUT101386,
  author="Zdeněk {Roubal} and Zoltán {Szabó} and Karel {Bartušek} and Miloslav {Steinbauer}",
  title="Measurement of concentration of air ions in the laboratories and cave",
  annote="A gerdien tube (GT) is instrumental towards the measurement of air ions concentration assuming that the volume of tested air has not been limited and the ions are continuously generated. In the time domain, it is possible to measure only ions of one polarity by means of a gerdien tube; then, following a certain interval, ions of the other polarity can be measured. Ionized air is sucked into the gerdien tube by a fan. A homogeneous electric field is set between the inner and the outer electrode. If an electric ion shows a negative electric charge and the collecting inner electrode has a positive electric potential, the ion is progressively attracted to the inner electrode. Provided that the ion impinges upon the electrode, it will induce an electric current that is measured by the help of a sensitive electrometric picoampermeter. The velocity of the ions motion in the electric field can be described by mobility k  [ m2*V-1*s-1].",
  booktitle="International Interdisciplinary PhD Workshop 2013, Proceedings",
  chapter="101386",
  howpublished="electronic, physical medium",
  year="2013",
  month="september",
  pages="182--186",
  type="conference paper"
}