Portable device for pesticide toxicity analysis

Originální název

Portable device for pesticide toxicity analysis

Anglický název

Portable device for pesticide toxicity analysis

Jazyk

en

Originální abstrakt

Fast detection of pesticide toxicity in field conditions is not fully solved yet. The new device for field measurement of organophosporous pesticides toxicity is described and first results are presented. The detection is based on the inhibition of acetylcholin esterase. The enzyme acetylcholin esterase is immobilized on a miniature electrochemical detector made by thick film technology forming together a biosensor. The dimensions of biosensor are 7,35 x 25,4 mm. The biosensor is placed in flow chamber which assures the reproducible hydrodynamic conditions near of active surface of the sensor. The analyzed sample is detected in the flow of supporting buffer. The biosensor detects the integral sample toxicity. It is destinated for detection of traces of pesticides in washout from leaves for direct measurment in rivers, ponds, vaste waters and drinking water sources. The preconcentration of sample is possible. The detection limit varies in wide range depending on the toxicity of pesticide. For example meviphos the detection limit is 8 . 10-6 mg/ml without preconcentration.

Anglický abstrakt

Fast detection of pesticide toxicity in field conditions is not fully solved yet. The new device for field measurement of organophosporous pesticides toxicity is described and first results are presented. The detection is based on the inhibition of acetylcholin esterase. The enzyme acetylcholin esterase is immobilized on a miniature electrochemical detector made by thick film technology forming together a biosensor. The dimensions of biosensor are 7,35 x 25,4 mm. The biosensor is placed in flow chamber which assures the reproducible hydrodynamic conditions near of active surface of the sensor. The analyzed sample is detected in the flow of supporting buffer. The biosensor detects the integral sample toxicity. It is destinated for detection of traces of pesticides in washout from leaves for direct measurment in rivers, ponds, vaste waters and drinking water sources. The preconcentration of sample is possible. The detection limit varies in wide range depending on the toxicity of pesticide. For example meviphos the detection limit is 8 . 10-6 mg/ml without preconcentration.

Dokumenty

BibTex


@inproceedings{BUT4944,
  author="Jan {Krejčí} and Radimír {Vrba} and Jiří {Háze} and Lukáš {Fujcik} and Michal {Skočdopole} and Luboš {Jakubka}",
  title="Portable device for pesticide toxicity analysis",
  annote="Fast detection of pesticide toxicity in field conditions is not fully solved yet. The new device for field measurement of organophosporous pesticides toxicity is described and first results are presented. The detection is based on the inhibition of acetylcholin esterase. The enzyme acetylcholin esterase is immobilized on a miniature electrochemical detector made by thick film technology forming together a biosensor. The dimensions of biosensor are         7,35 x 25,4 mm. The biosensor is placed in flow chamber which assures the reproducible hydrodynamic conditions near of active surface of the sensor. The analyzed sample is detected in the flow of supporting buffer.
The biosensor detects the integral sample toxicity. It is destinated for detection of traces of pesticides in washout from leaves for direct measurment in rivers, ponds, vaste waters and drinking water sources. The preconcentration of sample is possible.
The detection limit varies in wide range depending on the toxicity of pesticide. For example meviphos the detection limit is 8 . 10-6 mg/ml without preconcentration. 
",