Originální abstrakt

This work introduces a new method of detection of P, QRS and T components of ECG based on phasor transform (PT). PT converts each ECG sample into a phasor and therefore enhances changes in the ECG signal. Waves with small voltage are then easily detectable. The method has been tested on standard ECG databases MIT-BIH arrhythmia database, QT database and CSE database). The results of detection of P wave and QRS complex are better than results presented in previous works. The result of T wave detection is slightly worse than results of the most of previous works.

Anglický abstrakt

This work introduces a new method of detection of P, QRS and T components of ECG based on phasor transform (PT). PT converts each ECG sample into a phasor and therefore enhances changes in the ECG signal. Waves with small voltage are then easily detectable. The method has been tested on standard ECG databases MIT-BIH arrhythmia database, QT database and CSE database). The results of detection of P wave and QRS complex are better than results presented in previous works. The result of T wave detection is slightly worse than results of the most of previous works.

BibTex


@inproceedings{BUT127724,
  author="Lucie {Maršánová}",
  title="Detection of P, QRS and T Components of ECG Using Phasor Transform",
  annote="This work introduces a new method of detection of P, QRS and T components of ECG based on phasor transform (PT). PT converts each ECG sample into a phasor and therefore enhances changes in the ECG signal. Waves with small voltage are then easily detectable. The method has been tested on standard ECG databases MIT-BIH arrhythmia database, QT database and CSE database). The results of detection of P wave and QRS complex are better than results presented in previous works. The result of T wave detection is slightly worse than results of the most of previous works.",
  address="Vysoké učení technické v Brně",
  booktitle="Sborník příspěvků studentské konference Blansko 2016",
  chapter="127724",
  howpublished="online",
  institution="Vysoké učení technické v Brně",
  year="2016",
  month="august",
  pages="55--58",
  publisher="Vysoké učení technické v Brně",
  type="conference paper"
}