Publication detail

Analysis Of Gas Flow In The New System Of Apertures In The Secondary Electron Scintillation Detector For ESEM

MAXA, J. NEDĚLA, V. JIRÁK, J.

Original Title

Analysis Of Gas Flow In The New System Of Apertures In The Secondary Electron Scintillation Detector For ESEM

English Title

Analysis Of Gas Flow In The New System Of Apertures In The Secondary Electron Scintillation Detector For ESEM

Type

journal article - other

Language

en

Original Abstract

The work deals with analysis of gas flow in the detector, where the apertures A1 and A2 with one hole 0,6 mm in diameter are replaced by one aperture consisting of a system of parallelly distributed holes with a diameter of 0,1 mm

English abstract

The work deals with analysis of gas flow in the detector, where the apertures A1 and A2 with one hole 0,6 mm in diameter are replaced by one aperture consisting of a system of parallelly distributed holes with a diameter of 0,1 mm

Keywords

ESEM, gas flow, PLAs, SE detector

Released

27.07.2012

Location

USA DOI: 10.1017/S1431927612008173

Pages from

1264

Pages to

1265

Pages count

2

BibTex


@article{BUT93232,
  author="Jiří {Maxa} and Vilém {Neděla} and Josef {Jirák}",
  title="Analysis Of Gas Flow In The New System Of Apertures In The Secondary Electron Scintillation Detector For ESEM",
  annote="The work deals with analysis of gas flow in the detector, where the apertures A1 and A2 with one hole 0,6 mm in diameter are replaced by one aperture consisting of a system of parallelly distributed holes with a diameter of 0,1 mm",
  chapter="93232",
  number="Suppl 2",
  volume="2012(18)",
  year="2012",
  month="july",
  pages="1264--1265",
  type="journal article - other"
}