Publication detail

IMPACT HAMMER CALIBRATION

Original Title

IMPACT HAMMER CALIBRATION

Czech Title

IMPACT HAMMER CALIBRATION

Language

cs

Original Abstract

This paper describes the construction of the calibration station for the impact hammer Endevco 2302-5 and it explains the influence of the impact mass on the calibration process. Based on the previous bachelor thesis [1] it also explains the origin of the oscillations on the accelerome-ter output signal described there and it suggests the improvements of the calibration process.

Czech abstract

This paper describes the construction of the calibration station for the impact hammer Endevco 2302-5 and it explains the influence of the impact mass on the calibration process. Based on the previous bachelor thesis [1] it also explains the origin of the oscillations on the accelerome-ter output signal described there and it suggests the improvements of the calibration process.

Documents

BibTex


@inproceedings{BUT158672,
  author="Šimon {Bilík}",
  title="IMPACT HAMMER CALIBRATION",
  annote="This  paper  describes  the  construction  of  the  calibration  station  for  the  impact  hammer Endevco 2302-5 and it explains the influence of the impact mass on the calibration process. Based on the previous bachelor thesis [1] it also explains the origin of the oscillations on the accelerome-ter output signal described there and it suggests the improvements of the calibration process.",
  address="Vysoké učené Technické, Fakulta elektrotechniky a komunikačních technologií",
  booktitle="Proceedings of the 25nd Conference STUDENT EEICT 2019",
  chapter="158672",
  howpublished="electronic, physical medium",
  institution="Vysoké učené Technické, Fakulta elektrotechniky a komunikačních technologií",
  year="2019",
  month="april",
  pages="300--303",
  publisher="Vysoké učené Technické, Fakulta elektrotechniky a komunikačních technologií",
  type="conference paper"
}